Ultrasound is the term for mechanical waves with oscillations (frequency) above the audible range. Ultrasound is characterized by: The speed of sound (c) is an important property of the medium in which sound propagates. It is 340 m/s in air and about 1500 m/s in water. Tissue is similar to water, and has a speed of sound of around c = 1540 m/s. Since wavelength λ = c/f, the wavelength for medical ultrasound is between 0.15 mm and 0.75 mm. To compare, the wavelength for audible sound lies in the range of 2 cm to 20 m. Ultrasound of the body can be used for imaging from a distance of up to 20 cm. Within this scale, ultrasound has properties which remind one more of the properties of visible light than those typically associated with audible sound: Ultrasound is generated with the help of piezoelectric ceramics with electrodes on each side. This is called a transducer or probe (see figure 4).


Figure 4: Image of an ultrasound probe for cardiology. The cable comes out from the top left and is connected to the ultrasound machine. The front of the probe is about 20 x 15mm and a few individual rod elements are shown here. In practice, there are between 48 and 128 such elements.

The ceramics has the property that it compresses and expands in tact with applied electrical tension. This movement is transferred to the tissue. Conversely, mechanical pressure from the outside will generate an electrical potential. The probe can thus be used both as transmitter and receiver by switching quickly between transmitting and receiving mode. Imaging is based on puls-echo measurement. The same principle is used in radar and sonar. A short pulse is sent from the probe, propagates through the medium and returns a reflected pulse each time it meets a transition to a different tissue. The time it takes from sending the pulse until the echo is received by the probe gives a measure of the distance to the reflecting area.